Agua de refrigeración:
Caudal aprox. 5 L/h; temperatura máx. 30 °C; presión máx. 0,5 bar
Agua desmineralizada:
Does not apply
Aire de instrumentación:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Alimentación:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Calibración:
Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Calidad de muestra:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Caudal de muestra:
100 - 300 mL/min
Certificaciones:
CE, ETL certified to UL and CSA safety standards, UKCA
Conexión a tierra:
Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Contenido de la caja:
EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Fluoride Electrode and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
Dimensiones (A x A x P):
688 mm x 460 mm x 340 mm
Drenaje:
Atmospheric pressure, vented, min. Ø 32 mm
Garantía:
12 meses
Grado de protección:
IP44
Interferencias:
Metal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.
Límite de detección (LOD):
0.05 - 2.5 mg/L F-: 0.05 mg/L
0.1 - 5 mg/L F-: 0.1 mg/L
0.2 - 10 mg/L F-: 0.2 mg/L
Limpieza automática:
Sí; frecuencia programable a voluntad: cada 6 horas, cada 12 horas, diariamente, semanalmente
Material:
Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
Método de medición:
Discontinuous, direct measurement by combined Ion-Selective Electrode
Método de medida:
Ion-Selective Electrode
Número de corrientes de muestra:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parámetro:
Fluoride
Peso:
Max. 35 kg (77 lb)
Precisión:
Better than 2% full scale range for standard test solutions
Presión de muestra:
By external overflow vessel (open to atmospheric pressure)
Rango de medición:
0.05 - 2.5 mg/L F-
0.1 - 5 mg/L F-
0.2 - 10 mg/L F-
Requisitos de los reactivos:
Keep between 10 - 30 °C (50 - 86 °F)
Salidas analógicas:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Salidas digitales:
Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
RS485 communication: Profibus DP or Modbus RTU
Temperatura ambiente:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura de la muestra:
10 - 30 °C (50 - 86 °F)
Tiempo de ciclo:
Default: 5 minutes
Continuous: 5 minutes
Validación:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly